The JOFRA ITC series of dry-block calibrators has been designed for accurate true temperature calibration in the field. The ITC series features a modified dual-zone heating block from the ATC series. This means much better calibration uncertainties throughout a broad varieties of temperature sensors. The simple operation and the automatic functions have been carefully selected for ease of use. The ITC series is an efficient temperature calibrator without compromising quality.

 

  • Dual-zone heating block
  • Interchangeable inserts
  • Multi-hole insertion tubes
  • MVI Mains Variance Immunity
  • Automatic switch test
  • RS232 and JOFRACAL calibration software
  • Multi information display
  • Safety banana bushings
  • Auto stepping
  • Stability indicator
  • Count down timer (shows when unit is stable)
  • Programmable maximum temperature
  • Fast simulation/training mode
  • Re-calibration from keyboard

Model and temperature range
ITC-155 A..........-23 to 155°C (-9 to 311°F)
ITC-320 A..........50 to 320°C (122 to 608°F)
ITC-650 A..........50 to 650°C (122 to 1202°F)

 
 

 

See others dry-block calibrators:
JOFRA ATC Series

JOFRA CTC Series
JOFRA MTC Series
JOFRA ETC Series

     
     
  Well - optimized for accuracy - The ITC-320 and ITC-650 employ a modified dual-zone principle derived from the ATC series. This ensures good temperature homogeneity in the calibration zone. The ITC-320 and ITC-650 also use the same inserts as the ATC series. This means deep immersion depth and a diameter wide enough to leave room for also a reference probe or multiple sensors.
More about Dual-zone central-control
     
    Three models - The JOFRA ITC series consists of three models in different temperature ranges. All units offer excellent performance and feature the large, backlit LCD display panel, which is easy to read even in well lit areas. A rugged, slim-line, aluminium outer casing with die-cast top and bottom protects the ITC series of dry-blocks.

High flexibility - Interchangeable insertion tubes are used to match the diameter of your sensor-under-test. Also multihole inserts are available.
     

 
  Switch test - Fully automatic test of thermo switches. The calibrator will find the closing and the opening temperatures and calculate the hysteresis. This makes you save time and leaves you free to concentrate on other things while the test is running.

 
     
     

 
  Auto step - The calibrator will automatically change to the next programmed value. This allows the calibrator to be placed anywhere in the process while you monitor the output from the sensor-under-test in the control room.
     
    Easy maintenance - If adjustment of the calibrator is needed, use the built-in routine. No need for a computer or a screwdriver. "Calibration Date" is shown each time the instrument is switched on.
 
     

 
  MVI circuitry - The MVI technology removes the stability problem caused by an unstable mains supply. This is very often a problem in production environments with large electric motors, heating elements, etc. cycling on and off.
 
     

 
  Stability indicator - Clear audio and visual indication in the display when the calibrator is stable.
     
  Simplified calibration documentation - All ITC series calibrators are delivered with an RS232 serial interface and the JOFRACAL calibration software.
JOFRACAL software presents the means for an entirely automatic calibration of sensors and a semi-automatic calibration of the complete process loop through the use of a PC.

More about JOFRACAL.

Key specifications:
 
Spec's at 23°C (73°F) ambient
 
ITC-155 A
 
 
ITC-320 A
 
ITC-650 A
 
Min. temp. -23°C (-9°F)
 
50°C (122°F)
 
50°C (122°F)
 
Max. temp.
 
155°C (311°F)
 
320°C (608°F)
 
650°C (1202°F)
 
Dual-zone No
 
Yes
 
Yes
 
Stability
 
0.03°C (0.05°F)
 
0.03°C (0.05°F)
 
0.04°C (0.07°F)
 
Accuracy 0.25°C (0.45°F)
 
0.3°C (0.54°F)
 
0.5°C (0.9°F)
 
Immersion depth
 
160 mm (6.3 in)
 
150 mm (5.9 in)
 
150 mm (5.9 in)
 
Insert diameter
 
20 mm (3/4 in)
 
30 mm (1.2 in)
 
30 mm (1.2 in)
 

 
PRODUCT INFORMATION (Specification sheets)
You need the 'Acrobat Reader' to read PDF documents. Use this link to download the Adobe Acrobat Reader .

 

 

JOFRA ITC Series - - - - Dry-block - - - - Specification sheet in English


 

JOFRA ITC Series - - - - Dry-block - - - - Specification sheet in French

 


 

Insertion tubes for JOFRA temperature dry-blocks - - - - - Accessory sheet

 


 

User Manual JOFRA ITC-155/320/650 A

 

Reference Manual JOFRA ITC-155//320/650 A
 

Request more information about this product
 



See more

JOFRA DRY-BLOCK CALIBRATORS

Ametek Temperature Calibrators

 


     Subscribe to our product specials newsletter.

  Terms and Conditions :: Privacy Policy

© 2009 Digital Measurement Metrology, Inc.
Toronto, Ontario, CANADA